Labeling table (blood, plasma and Platelet )+4C ,+22C , -25C
Application :
It is used in blood transfusion organizations
for labeling blood and plasma bags
Technical Specifications
• SECOP compressor
• Eliwell controller : 3 pcs with NTC sensor
• 3 different temperatures : +4°C for releasing blood ,
+22°C for releasing platelet ,-25°C for releasing plasma
• Blood release temperature tolerance : ±2°C
• Plasma release temperature tolerance : ±3°C
• Platelet release temperature tolerance : ± 2°C
• Recorder and controller accuracy : 0.1°C
• Internal chamber : stainless steel 314
• External body : Galvanized steel with electrostatic coating
• Polyurethane insulation
• European digital controller : 3 pcs
• Temperature recorder with one PT100 sensors for recording temperature , date and time every 5 minutes (last two years are available
in the memory) : Displaying and drawing graph, table and alarm for 1channels, USB
port for downloading data to flash memory in excel file , RS-485 (Modbus-RTU) for temperature monitoring network , Backup battery to
keep the recorder running in case of power failure for 4 hours (only for recording data and sending alarm)
• Option: SMS center can be installed on it for sending alarm message and temperature
• Volume :180 L
• Power input :700VA